14423-2 / 14423-4 / 14423-8
ULTRASOUND POCKET DOPPLER
USER MANUAL
14423-INS-LAB-RevD11
Important: Do not use the Pocket Doppler
without rst reading and understanding this
manual! Save this manual for future use.
FederalLaw(USA)restrictsthisdevicetosale
byorontheorderofaphysician.Thisproduct
shouldnotbeusedunlesstheoperatorhas
beeninstructedbyaqualiedhealthcarepro-
fessional.
Vue de la page 0
1 2 3 4 5 6 ... 43 44

Résumé du contenu

Page 1 - USER MANUAL

14423-2 / 14423-4 / 14423-8ULTRASOUND POCKET DOPPLERUSER MANUAL14423-INS-LAB-RevD11Important: Do not use the Pocket Doppler without rst reading and u

Page 2 - CONTENTS

14423-INS-LAB-RevD11 103 HANDLINGGFHEALTHPRODUCTS,INC.FREIGHTPOLICYForyourprotection,readcarefullyThe carrier accepted this merchandise “in

Page 3 - 1 INTRODUCTION

14423-INS-LAB-RevD11 11InspectionCheck the Pocket Doppler for nicks, dents, scratches, me-chanical or other damage. Check all the cables and acces-so

Page 4 - OPTIONALACCESSORIES

14423-INS-LAB-RevD11 124 ULTRASOUNDPOCKETDOPPLERANDACCESSORIESAPPEARANCEANDFEATURESThe following gures illustrate the Pocket Doppler with the

Page 5 - 2 SAFETYINFORMATION

14423-INS-LAB-RevD11 13PowerButtonTurns the Ultrasound Pocket Doppler ON or OFF ProbeSocketThe probe connects here to the Ultrasound Pocket Doppler

Page 6 - 14423-INS-LAB-RevD11 6

14423-INS-LAB-RevD11 14RearPanelFeature FunctionPocketClipAttachmentAttaches Pocket Doppler to pocketBatteryCompartmentOpens to install/replace b

Page 7 - 14423-INS-LAB-RevD11 7

14423-INS-LAB-RevD11 15TopPanelFeature FunctionSocket Name Picture FunctionCharge SocketNot usedEarphone SocketConnect earphoneSignal Interface1 GND

Page 8 - 14423-INS-LAB-RevD11 8

14423-INS-LAB-RevD11 16LeftPanelFeature FunctionVolumeControlAdjust volumeIncrease volume: Rotate volume knob clockwiseDecrease volume: Rotate volu

Page 9 - 14423-INS-LAB-RevD11 9

14423-INS-LAB-RevD11 17POCKETDOPPLERWATER-RESISTANTPROBESThe 2.0 MHz, 4.0 MHz, and 8.0 MHz water-resistant probes connect to the main unit of the

Page 10 - 3 HANDLING

14423-INS-LAB-RevD11 185 SETUPINSTALLINGORREPLACINGTHEBATTERYWARNING:EnsurethatPocketDopplerpowerisoffbeforereplacingthebattery.WAR

Page 11 - 14423-INS-LAB-RevD11 11

14423-INS-LAB-RevD11 19s NOTICE:Ifthebatteryterminalsbecomedirty,wipethemwithaclean,dryclothbeforeusingthePocketDoppler.s NOTICE:

Page 12 - ANDACCESSORIES

14423-INS-LAB-RevD11 2CONTENTS1 INTRODUCTION...3INTENDEDUSEOFTHISDEVICE.

Page 13 - Jack Denition

14423-INS-LAB-RevD11 206 OPERATIONPROBEOPERATIONRemovingtheProbefromthePocketDopplers NOTICE:Donotdragordroptheprobe.Donotdis-conn

Page 14 - 14423-INS-LAB-RevD11 14

14423-INS-LAB-RevD11 21SwappingProbess NOTICE:Donotdragordroptheprobeortheprobeconnector.The Pocket Doppler is shipped with one probe co

Page 15 - 14423-INS-LAB-RevD11 15

14423-INS-LAB-RevD11 22FETALHEART(FH)MONITORINGFHMonitoringwith2.0MHzWater-ResistantProbeRefer to picture at right. The au-dio fetal heart b

Page 16 - 14423-INS-LAB-RevD11 16

14423-INS-LAB-RevD11 23VASCULARMONITORINGWITH4.0OR8.0MHZPROBE(OPTIONAL)The 4.0 MHz and 8.0 MHz probes can be used to perform vascular monitor

Page 17 - 14423-INS-LAB-RevD11 17

14423-INS-LAB-RevD11 241. Apply a liberal amount of gel on the site to be exam-ined.2. Place the probe so that the head is at least at 45º to the v

Page 18 - 5 SETUP

14423-INS-LAB-RevD11 257 MAINTENANCEWARNING:Failureonthepartoftheresponsiblefacil-ityemployingtheuseofthisequipmenttoimplementasa

Page 19 - 14423-INS-LAB-RevD11 19

14423-INS-LAB-RevD11 26s NOTICE:KeepthePocketDoppleranditsenviron-mentcleananddust-free.Protectthedevicefromvi-bration,corrosivemed

Page 20 - 6 OPERATION

14423-INS-LAB-RevD11 27CleaningthePocketDopplerProbes NOTICE:Donotallowliquidtoentertheultrasoundsystem.s NOTICE:Donotpourliquids

Page 21 - TURNINGONTHEPOCKETDOPPLER

14423-INS-LAB-RevD11 28DISINFECTIONDisinfectingthePocketDopplerProbes NOTICE:Donotallowanyliquidtoentertheprobesocketwhiledisinfecti

Page 22 - FETALHEART(FH)MONITORING

14423-INS-LAB-RevD11 298 SPECIFICATIONSModelNumberandName14423-2 Ultrasonic Pocket Doppler with 2.0 MHz probe14423-4 Ultrasonic Pocket Doppler wi

Page 23 - PROBE(OPTIONAL)

14423-INS-LAB-RevD11 31 INTRODUCTIONThis user manual contains important information and safety precautions for the Grafco Ultrasound Pocket Dop-ple

Page 24 - REPLACINGTHEBATTERY

14423-INS-LAB-RevD11 30PhysicalCharacteristicsSize (Width x Depth x Height)3.35 x 1.26 x 5.43 inches (85 x 32 x 138 mm)Weight (including a battery a

Page 25 - 7 MAINTENANCE

14423-INS-LAB-RevD11 31UltrasoundNominal Frequency2.0 MHz Water-Resistant Probe2.0 MHz4.0 MHz Water-Resistant Vascular Probe4.0 MHz8.0 MHz Water-Resi

Page 26 - CLEANING

14423-INS-LAB-RevD11 329 LIMITEDWARRANTYGF Health Products Inc. warrants this product to be free from manufacturing defects in material and wo

Page 27 - 14423-INS-LAB-RevD11 27

14423-INS-LAB-RevD11 33purchaser’s cost, for warranty inspection using the serial number as identication (or, if the product is not serialize

Page 28 - DISINFECTION

14423-INS-LAB-RevD11 34This warranty gives you specic legal rights and you may also have other legal rights which vary from state to state (prov-inc

Page 29 - 8 SPECIFICATIONS

14423-INS-LAB-RevD11 3510APPENDIXA,ECDECLARATIONOFCONFORMITYECDeclarationofConformityManufactured for GF Health Products, Inc., 2935 Northea

Page 30 - 14423-INS-LAB-RevD11 30

14423-INS-LAB-RevD11 3611APPENDIXB,EMCINFORMATIONGUIDANCEANDMANUFACTURER’SDECLARATIONElectromagneticEmissions—forallEquipmentandSystemsGu

Page 31 - Ultrasound

14423-INS-LAB-RevD11 37ElectromagneticImmunity—forallEquipmentandSystems thatarenotLife-SupportingGuidanceandmanufacturer’sdeclaration—ele

Page 32 - 9 LIMITEDWARRANTY

14423-INS-LAB-RevD11 38Info 1: At 80 MHz and 800 MHz, the higher frequency range applies.Info 2: These guidelines may not apply in all situations.

Page 33 - 14423-INS-LAB-RevD11 33

14423-INS-LAB-RevD11 39For transmitters rated at a maximum output power not listed previously, the recommended separation distances in meters (m) can

Page 34 - 14423-INS-LAB-RevD11 34

14423-INS-LAB-RevD11 4OPTIONALACCESSORIESThe following accessories for use with the Grafco Ultrason-ic Pocket Doppler are available from your GF Hea

Page 35 - CONFORMITY

14423-INS-LAB-RevD11 4012APPENDIXC,OVERALLSENSITIVITYOverallSensitivity(2MHzProbe)Diameter of Target Reector (mm)Distance (d) (mm)Reflection

Page 36 - DECLARATION

14423-INS-LAB-RevD11 41OverallSensitivity(2MHzProbe)Diameter of Target Reector (mm)Vn (r.m.s.) mVVn (r.m.s.) mVdBOverall Sensitivity (S=A(d)+B+C

Page 37 - 14423-INS-LAB-RevD11 37

14423-INS-LAB-RevD11 4213INDEXAAccessories, optional 4Accessory equipment, certica-tion of 6Address, Graham-Field 34Appendix A, EC Declaration o

Page 38 - 14423-INS-LAB-RevD11 38

14423-INS-LAB-RevD11 43TTelephone number, Graham-Field 34Top panel features and illustra-tion 15UUser manual, contents of 3VVolume control 16WWAR

Page 39 - 14423-INS-LAB-RevD11 39

USA Corporate Headquarters:GF Health Products, Inc. 2935 Northeast Parkway Atlanta, Georgia 30360telephone: 770-368-4700www.grahamfield.com14423-INS-

Page 40 - 14423-INS-LAB-RevD11 40

14423-INS-LAB-RevD11 52 SAFETYINFORMATIONSafetyGuidanceThis unit has internally powered equipment, and the degree of shock protection is type B Ty

Page 41 - 14423-INS-LAB-RevD11 41

14423-INS-LAB-RevD11 6WARNINGSWARNING:FederalLaw(USA)restrictsthisdevicetosalebyorontheorderofaphysician.Thisproductshouldnotbe

Page 42 - 13INDEX

14423-INS-LAB-RevD11 7WARNING:DonottheusePocketDopplerifthebat-terycoverisopen.WARNING:UseonlythoseprobesprovidedbyGFHealthPro

Page 43 - 14423-INS-LAB-RevD11 43

14423-INS-LAB-RevD11 8NOTICESs NOTICE:ThemainunitoftheGrafcoPocketDopplerisdesignedforcontinuousnormaloperation.Donotimmerseorsub

Page 44

14423-INS-LAB-RevD11 9SYMBOLSAttention: Refer to accompanying documents (this manual). This item is compliant with Medical Device Directive 93/42/EEC